Intercomparison of Gamma Cell 220 Irradiator Facilities and Dr. Mirzan T Razzak Gamma Irradiators Using Harwell Dosimeters
DOI: http://dx.doi.org/10.17146/jfn.2021.15.1.6284
Sari
The gamma irradiator is a multi-purpose facility that possibly used to preserve food, sterilize medical equipment, and conduct genetic engineering and polymerization processes, during which the absorbed dose of the product is critical. The standardization of product quality assurance was regulated by the IAEA Technical Document Number 409 considering Dosimetry for Food Irradiation and ISO 14470 and 11137-3 on Food Irradiation, as well as the Guidance on Dosimetric Aspects of Development, Validation, and Routine Control, respectively. The absorbed dose was influenced by the movement of the product to the source, its position, the amount of radioactive activity in the facility, and the dose rate in the irradiation room. The dosimeter performance test and quality assurance of the system were conducted using the Facility Intercomparison Technique which tested the dosimeter (measuring instrument) at 2 different facilities to determine the performance of the measuring instrument.. In this study, 2 irradiation facilities were tested using a Harwell routine dosimeter in the dose range of 1 kGy to 30 kGy and 20 dose points. The results showed that the highest deviation reached 19% and 21% at the Gamma Cell 220 and the Dr. Mirzan T Razzak Gamma irradiator facilities. This elevated the performance of the dosimeters to determine the precision accuracy of the dose-measuring instrument.
Teks Lengkap:
PDF (English)Referensi
- IAEA, 2015, Technical Report Series no.48: Manual of Good Practice in Food Irradiation, Vienna.
- INTERNATIONAL STANDART, 2011, ISO 14470: Food Irradiation-Requirement for the Development, Validation and Ruotine Control of the Process of Irradiation Using Ionizing Radiation for the Treatment of Food, Geneva: Swiss.
- INTERNATIONAL STANDART, 2006, ISO 11137: Requirement for Development, Validation and Routine Control of a Sterilization Process for Medical Devices, Geneva: Swiss.
- INTERNATIONAL STANDART, 2013, ISO 11137: Establishing of Sterilization Dose, Geneva: Swiss.
- INTERNATIONAL STANDART, 2017, ISO 11137: Guidance of Dosimetric Aspects of Development, Validation and Routine Control, Geneva: Swiss.
- R. G. Aghoyeh, H. Khalafi, “Design of dual-column water purification system for industrial gamma irradiator based of PUROLITE resins,” Journal Annals of Nuclear Energy Series 69, pp. 90-96, 2014.
- K. Manai, B. Askri, A. Loussaief, A. Trabelsi, “Evaluation Using GEANT4 of transit dose in Tunisian Gamma Irradiator for Insect Sterilization,” Journal Applied Radiation and Isotopes, pp. 701-707, 2007.
- V. Drndarevic, “Control of Gamma Irradiation Facility with Improved Safety System,” Journal of Nuclear Science and Technology, vol. 45(4), pp. 361-367, 2012.
- K. Elhamdi, K. Manai, M. Bhar, “Dose Calculation Using Tchebyshev Wavelets with Buildup Correction in The Tunisian Gamma Irradiator,” Journal Radiation Physics and Chemistry, vol. 150, pp. 46-50, 2018.
- M. D. F. Mortuza, L. Lapore, K. Khedkar, S. Thangam, A. Nahar, H. M. Jamil, L. Bandi, Md. K. Alam, “Commisioning Dosimetry and In Situ Dose Mapping of Semi Industrial Cobalt-60 Gamma Irradiation Facility Using Fricke and Ceric Cerous Dosimetry System and Comparison with Monte Carlo Simulation Data,” Journal Radiation Physics, and Chemistry, 2017.
- N. S. Pamungkas, R. B. Cahyono, D. Swantomo, “Penilaian Safety Integrity Level Iradiator Gamma Kategori-IV Pada Kegagalan Sistem Crane Pengontrol Sumber Zat Radioaktif,” Jurnal Forum Nuklir, vol. 12(1), pp. 1-6, 2018.
- O. Nugroho, S. Putra, N. A. Kundari, “Prarancang Unit Pengolahan Limbah Cair Pabrik Jeans Dengan Teknologi Iradiasi Gamma,” Jurnal Forum Nuklir, vol. 13(2), pp. 63-72, 2019.
Refbacks
- Saat ini tidak ada refbacks.