MODIFIKASI METODA ASTM UNTUK ANALISIS URANIUM DENGAN KONSENTRASI 1 gU/L MENGGUNAKAN TITROPROSESOR

Sigit Sigit, Noor Yudhi, Rahma Yanda

DOI: http://dx.doi.org/10.17146/urania.2013.19.3.1801

Sari


ABSTRAK

MODIFIKASI METODA ASTM UNTUK ANALISIS URANIUM DENGAN KONSENTRASI 1 gU/L MENGGUNAKAN TITROPROSESOR. Telah dilakukan penentuan uranium dalam larutan radioaktif secara titrasi potensiometrik menggunakan titroprosesor dengan memodifikasi metode ASTM C 799-83 yaitu pengurangan volume pereaksi menjadi 10, 15, 20 dan 30% dari masing-masing volume pereaksi yang digunakan dalam ASTM yaitu asam sulfamat, asam pospat, ferro sulfat, ammonium molibdat dan vanadium sulfat. Di samping itu juga dipelajari variasi keasaman 1, 2, 3 dan 4N serta konsentrasi uranium dalam umpan 1, 2, 3 dan 4 g/L. Tujuan penelitian adalah memperoleh kondisi relatif baik untuk analisis uranium dalam larutan yang mengandung uranium sekitar 1 g/L yang kemudian dapat diterapkan untuk analisis uranium dalam limbah cair radioaktif. Sampel dimasukan ke dalam gelas piala, lalu elektroda platina dimasukkan hingga tercelup dan titrasi potensiometrik dimulai dengan menjalankan titroprosesor. Hasil percobaan menunjukkan bahwa larutan yang mengandung uranium dengan konsentrasi 1 gU/L dapat dianalisis dengan modifikasi volume pereaksi 10% dari volume pereaksi pada ASTM dan memberikan presisi 3,02%, RSD 0,0302 dan akurasi 95,11%, sedangkan untuk variasi keasaman, diperoleh hasil relatif baik yaitu keasaman 4N yang ditunjukkan oleh presisi dan penyimpangan yang paling kecil dengan akurasi >95% dan dapat digunakan untuk analisis uranium pada konsentrasi 1-4 g/L dengan presisi 1,33 – 3,17 dan akurasi ≥99,99%. Pada analisis limbah efluen (LE) cair radioaktif dengan kode LE-15 dan LE-3 menggunakan modifikasi metode ASTM diperoleh konsentrasi uranium masing-masing 0,0558 g/L dan 0,6139 g/L.

Kata kunci: modifikasi metode ASTM, penentuan uranium, limbah radioaktif cair,  titroprosesor.

 

ABSTRACT

MODIFICATION OF ASTM METHOD FOR URANIUM ANALYSIS WITH CONCENTRATION 1 gU/L USING TITROPROCESSOR. Determination of uranium concentration in radioactive solution has been done by modified potentiometric method of ASTM C 799-83. The modification includes reagent volume reduction to 10%, 15%, 20% and 30% of reagent volume used in the ASTM, i.e. sulfamic acid, phosphate acid, ferrous sulfate, ammonium molybdate and vanadium sulfate. This research also studied variation in the acidity of 1, 2, 3, and 4N and uranium concentration in the feed of 1, 2, 3 and 4 g/L. The research objective is to determine the relatively good condition for the analysis of uranium solution containing approximately 1 g/L uranium. This modified method may be applied to the analysis of uranium in radioactive liquid waste. Samples were loaded into a glass cup with a platinum electrode immersed, and potentiometric titration began by running titroprosesor. The results show that a solution containing uranium with a concentration of 1 g/L can be analyzed with the modifified reagent volume to 10% of the reagent volume according to the ASTM with 3.02% precision, 0.0302 RSD and 95.11% accuracy. The variation of acidity gave good result at a relative acidity of 4N, which is shown by the precision and the smallest deviation with an accuracy of > 95%. This condition can be used for the analysis of uranium at a concentration of 1-4 gU/L with a precision of 1.33 to 3.17 and an accuracy of ≥ 99,99%. Analysis of radioactive liquid waste labeled with LE-15 and LE-3 by this modified ASTM method gives a value of uranium concentration of 0,0558 g/L and 0,6139 g/L respectively.

Keywords: ASTM method modification, uranium determination, liquid radioactive waste,      titroprocessor.

Teks Lengkap:

PDF

Refbacks

  • Saat ini tidak ada refbacks.


Penerbit: Pusat Riset Teknologi Daur Bahan Bakar Nuklir dan Limbah Radioaktif 

Diindeks oleh:

   
     
      

p-ISSN 0852-4777 | e-ISSN 2528-0473